NEW CALIBRATOR WITH POINTS DISTRIBUTED CONICAL HELICALLY FOR ONLINE CALIBRATION OF C-ARM

New Calibrator with Points Distributed Conical Helically for Online Calibration of C-Arm

New Calibrator with Points Distributed Conical Helically for Online Calibration of C-Arm

Blog Article

To improve the ngetikin accuracy of calibration of C-arm, and overcome the space limitation in surgery, we proposed a new calibrator for online calibration of C-arm.After the image rectification by a polynomial fitting-based global correction method, the C-arm was assumed as an ideal pinhole model.The relationships between two kinds of spatial calibration errors and the distribution of fiducial points were studied: the performance of FRE (Fiducial Registration Error) and TRE (Target Registration Error) were not consistent, but both were best at the 12 marked points; the TRE decreased with the increase of the uniformity of calibration points distribution, and with the decrease of the distance between the target point and the center of calibration points.A calibrator with 12 fiducial points conical helically distributed, which could be placed on the knee, was an attractive option.

A total of 10 experiments on C-arm calibration accuracy were conducted and the mean value of mapping error was 0.41 mm.We designed an ACL reconstruction navigation system deva curl arc angel and carried out specimen experiments on 4 pairs of dry femur and tibia.The mean accuracy of navigation system was 0.

85 mm, which is important to the tunnel positioning for ACL reconstruction.